Understanding Fluorescence
Fig. 1
Source: http://www.lookfordiagnosis.com/mesh_info.php?term=Spectrometry%2C+Fluorescence&lang=1
The figure
above (Fig. 1) shows the typical instrumentation used to make fluorescence
measurements. The light source could be
a xenon arc lamp or a mercury vapor lamp.
These produce a wide range of wavelengths, most importantly in the UV
region. The monochromator,
a diffraction grating, splits the incoming collimated light into a
spectrum. The desired wavelength is
selected using a narrow slit. The “excitation”
light hits the sample and the sample emits light at longer wavelengths. This fluorescent light must be collimated and
can be analyzed using a second monochromator.
As shown
below (Fig. 2) for rhodamine (a widely used fluorescent label by which
molecules can be tagged), two types of spectra can be generated: an excitation
spectrum and an emission spectrum. The
excitation spectrum shows the wavelengths of light that can be used to excite a
molecule to produce fluorescent light.
Question: How would one go about
collecting an excitation spectrum using the instrumentation illustrated in Fig.
1?
The other
spectrum is the emission spectrum. It
shows the range of wavelengths of fluorescent light emitted by the fluorescent
molecule.
Question: How would one go about
collecting an emission spectrum using the instrumentation illustrated in Fig. 1?
Fig. 2
Source: http://elchem.kaist.ac.kr/vt/chem-ed/spec/molec/mol-fluo.htm
Looking at the two spectra, you notice that both consist of a range of wavelengths rather than just one wavelength. There are a variety of reasons for the range of wavelengths.
Task: Looking at Fig. 3, propose one source for the range of wavelengths seen in each of the spectra?
Fig. 3 Source: https://web.nmsu.edu/~kburke/Instrumentation/fluorescence_Std_4.html
Considering Fig. 3:
Question: Does the
shape of the excitation spectrum depend on the wavelength used to monitor the
fluorescence? Why?
Question: Does the
shape of the emission spectrum depend on the wavelength used to excite the
fluorescent molecule? Why?
Fig. 4
Source: http://biomedicaloptics.spiedigitallibrary.org/article.aspx?articleid=2546046
If the fluorescence is measured as a function of concentration (Fig. 4) of the fluorescent molecule, the fluorescence is often non-linear, saturating and declining at high concentrations.
Question: What might be the cause of this non-linear
behavior? (look
at Fig. 2 for a clue)
Photobleaching
Colors of objects often fade when exposed to sunlight: e.g. bumper-stickers, clothes, printed photographs.
Question: What causes
this fading? Explain it at the molecular
level.
This fading is photobleaching.
Question: Which colors are more likely to photobleach, everything else being equal? Why?
Question: Would
fluorescent molecules be more sensitive to photobleaching
than colored molecules? Why?
Quenching and Quantum Yield
When an electron in a fluorescent molecule absorbs a photon and moves to an excited state, it does not always emit a photon to return to the ground state. The quantum yield is the ratio of photons emitted divided by the photons absorbed.
Question: How could
the quantum yield be measured? Is it just the number of fluorescence photons
measured by the detector in Fig. 1 divided by the number of photons sent to the
cuvette in Fig. 1?
Molecules in the environment can reduce the fluorescence measured. This process is called quenching. For example, both water and oxygen dissolved in water can quench the fluorescence but some specific molecules are much more effective. These are referred to as quenchers.
Question: How might
quenchers work at the molecular level?
FRAP: Flourescence Recovery After Photobleaching or FRAP is a method developed to measure the lateral mobility of the constituents of a membrane. Look it up in Wikipedia.
Question: How does
this method work?
Question: How can the same wavelength of light be used
both to measure fluorescence and to “bleach” the fluorescent molecule?
Question: If a 4micrometer
diameter bleached spot is made in the plasma membrane of a cell, in what time
scale would the recovery take place for a protein?...for a phospholipid
molecule? (assume
they move freely)